Safety during operation – VACUUBRAND RZ 9 User Manual

Page 12

Advertising
background image

page 12 of 49

• Check the inlet and outlet of the pump and its oil condi-

tion, if there is a danger of forming

deposits inside the

pump, e.g., in the pump unit. Inspect the pump regu-

larly and clean if necessary.

• Install a cold trap in front of the pump if pumping ag-

gressive or corrosive gases or vapors.

Consider interactions and chemical reactions of

the pumped media. Ensure that the materials of the

pump’s wetted parts are compatible with the pumped

substances, see section “Technical data”, pg. 17.

When changing the substances pumped, we recom-

mend purging the pump with air or inert gas prior to

changing the pumped media. Purging the pump will

pump out residues and it will reduce the possibility of

reactions of the pumped substances with each other

and with the pump’s materials.

Safety during operation

Adopt suitable measures to prevent the release of dan-

gerous, toxic, explosive, corrosive, noxious or pollut-

ing fluids, vapors and gases. To prevent any emission

of such substances from the pump outlet, install an

appropriate collecting and disposal system and take

protective action for pump and environment.

You must take suitable precautions to prevent any for-

mation of explosive mixtures in the expansion cham-

ber, in the oil reservoir, or at the outlet of the pump. In

such circumstances, mechanically generated sparks,

hot surfaces or static electricity may ignite these mix-

tures. Use inert gas for gas ballast or venting, if neces-

sary.

Drain appropriately or otherwise remove any poten-

tially explosive mixtures at the outlet of the pump, or

dilute them to non-explosive concentrations.