inlet or outlet.

• Pumping at high inlet pressure may lead to overpres-

sure at the gas ballast valve. Pumped gases or con-

densate might be expelled if the valve is open. If an in-

ert gas supply is connected to the gas ballast, ensure

that its inlet pipeline is not contaminated.

• Check oil level of the rotary vane pump (through sight

glass) every time before starting the pump.

• Pay attention to the safety symbol ”hot surfaces” on

the equipment.

During operation the surface temperature of the pumps

may exceed 176°F (80°C). Make sure not to touch the

pumps accidentally.

Hot parts may cause burns if touched. Adopt suitable

measures to prevent any danger arising from hot sur-

faces or electric sparks. Ensure that hot surfaces of

the pump do not cause burns. Provide a suitable con-

tact guard if necessary.

• A power failure may cause accidental ventilation of the

pump or the vacuum system, if the manual gas ballast

valve (RE 2.5 / RZ 2.5 / RE 6 / RZ 6 / RE 9 / RZ 9) is

open. If this constitutes a potential source of danger,

take appropriate safety measures, e.g., install a sole-

noid operated gas ballast valve.

Prevent the backpressure of gases and the backflow of

condensates at the outlet.

Provide appropriate protective measures to allow for the

possibility of failure and

malfunction. The protective mea-

sures must also allow for the requirements of the respec-

tive application.

Single-phase motors are protected by a self-hold ther-

mal cutout in the winding.

Three-phase motors are protected by a circuit-breaker

isolating all poles in case of overload.