Notice – VACUUBRAND RZ 9 User Manual

Page 27

Advertising
background image

page 27 of 49

• A power failure may cause accidental ventilation of the

pump or the vacuum system, especially if the manual

gas ballast valve (RE 2.5 / RZ 2.5 / RE 6 / RZ 6 / RE 9

/ RZ 9) is open. If this constitutes a potential source of

danger, take appropriate safety measures, e.g., install

a solenoid operated gas ballast valve (see “Accesso-

ries”, pg. 36).

• Check the power source and the pump’s rating plate

to be sure that the power source and the equipment

match in voltage, phase, and frequency.

Pump with dual-voltage motor: Check that the volt-

age selection switch at the terminal box is positioned

correctly. Check every time before starting the pump.

Note: If the pump is switched on with wrong voltage

selection, the motor may be damaged!

Change the selection at the voltage selection

switch only, if the pump is unplugged from the

power source.

Keep a minimum distance of 2 in (5 cm) between the cool-

ing fan and surrounding items (e.g., housing, walls, etc.),

or else install an external automatic ventilation system.

Use connecting hoses with large diameter and keep them

as short as possible to avoid flow losses. Locate the pump

as closely as possible to the application.

Always install outlet tubing descending from the pump or

provide other measures (e.g., install an oil mist separator

at the outlet) to avoid backflow of condensate towards the

pump.

Use a suitable valve to isolate the pump from the vacuum

application. This is to allow the pump to warm up before

pumping condensable vapors and to clean the pump after

use before it is switched off.

When assembling, ensure

vacuum-tightness. After as-

sembly, check the whole system for leaks.

Secure hose connections at the pump appropriately, e.g.,

NOTICE

Advertising
This manual is related to the following products: