During operation – VACUUBRAND RZ 9 User Manual

Page 28

Advertising
background image

page 28 of 49

with hose clamps, to protect against accidental detach-

ment.

During operation

Vent and dispose of

potentially dangerous gases or

vapors at the outlet of the pump appropriately.

+

Due to the high compression ratio, the pump might

generate overpressure at the outlet. Check pressure

compatibility with system components (e.g., exhaust

tubing or exhaust valve) at the outlet. Ensure that the

pump outlet is neither blocked nor restricted.

+

Maximum ambient temperature: 104 °F (40 °C)

Check the maximum temperatures, if installing the

pump in a cabinet or a housing. Make sure ventilation

is adequate, especially if the ambient temperature is

elevated.

• If the pump is installed at an altitude of more than 6500

ft (2000 m) above mean sea level, check compatibility

with applicable safety requirements, and adopt suit-

able measures. There is a risk of the motor overheat-

ing due to insufficient cooling.

• Check compatibility with the

maximally permitted

pressure at outlet and the maximum pressure dif-

ference between inlet and outlet ports.

• Check the

oil level of the rotary vane pump every time

before starting the pump, however at least once a week.

Check the oil level more frequently if high amounts of

gas or vapor are pumped.

• If pumping aggressive, corrosive or otherwise danger-

ous gases and vapors, take appropriate measures to

protect personnel, pump, and environment. Use ap-