• Check the power source and the pump’s rating plate

to be sure that the power source and the equipment

match in voltage, phase, and frequency.

Pump with dual-voltage motor: Check that the volt-

age selection switch is positioned correctly. Check ev-

ery time before starting the pump.

Note: If the pump is switched on with wrong voltage

selection, the motor may be damaged!

Change the selection at the voltage selection

switch only, if the pump is unplugged from the

power source.

Direction of rotation of pumps with three-phase mo-

tors: Pumps with three-phase motors wired ready for

connection are supplied with a power plug (CEE). The

electrical wiring is laid out for a right-handed rotary

field.

Checking the direction of rotation:

Switch on pump for a few seconds and check if the

pump aspirates at inlet. If no suction occurs switch off

pump immediately. The three phase socket has to be

checked and if necessary corrected by an electrician.

• Use only oil of the recommended type. Other oils or

operating fluids may cause damage of the pump or

danger.

• Use

special oils (see “Oil change”, pg. 32) for the

for the

rotary vane pump, if, e.g.,

- acid vapors are pumped.

- extremely high aging stability is required.

- oxygen or other strong oxidizing agents are pumped.

• Take adequate precautions when handling pump flu-

ids, lubricants, and solvents. Use appropriate protec-

tive clothing, safety goggles and protective gloves to

avoid excessive contact with the skin and possible skin

irritations (including dermatitis).

• Comply with all relevant statutory requirements and

regulations concerning the handling, storage and dis-

posal of oil.