propriate equipment such as cold trap, separator, oil

separator, full flow oil filter, shut-off valve (see “Acces-

sories”, pg. 36) as well as special oil.

The

ambient temperature should be at least 54°F (12

°C). Otherwise the pump may not start because of the

high oil viscosity at low temperature.

Pumping down can be started at any pressure at the inlet

below atmospheric pressure.

Do not start the pump, if the

pressure at the outlet port

exceeds 16.0 psi (1.1 bar) absolute.

Continuous operation is possible at any pressure below

atmospheric pressure. For pumps RE/RZ 16 with oil mist

separator, continuous operation is only possible below 75

Torr (100 mbar).

Oil consumption will increase at inlet pressures above 75

Torr (100 mbar). Check oil level more frequently.

Prevent internal condensation, transfer of liquids or dust.

The pump unit will be damaged if liquids are pumped in

significant amounts. Check the pump regularly for exter-

Check the pump regularly for exter-

Check the pump regularly for exter-

nal soiling and deposits. Clean the pump if necessary to

avoid an increase of the pump’s operating temperature.
Single-phase motors are protected by a self-hold ther-

mal cutout in the winding.

Three-phase motors are protected by a circuit-breaker

isolating all poles in case of overload.

Note: Only a manual reset is possible. Switch off the

pump and disconnect the electrical power cord. Identify

and eliminate the cause of failure. Wait approximately five

minutes before restarting the pump.

Avoid high heat supply (e.g., due to hot process gases).

See “Technical data”, pg. 17, for maximum permitted

gas and ambient temperatures.

A warm up period (approximately 30 min.) is required to

ensure that the rated ultimate vacuum and pumping speed

NOTICE