Oil change procedure – VACUUBRAND RZ 9 User Manual

Page 33

Advertising
background image

page 33 of 49

(8)

(9)

(7)

(1)

(10)

Under normal operating conditions:

- Check oil level every time before starting the pump.

- Carry out an oil change once aging of the oil is indi-

Carry out an oil change once aging of the oil is indi-

ging of the oil is indi-

cated by darker color compared to new oil.

- Carry out oil changes at least once a year.

- Dispose of the used oil, which may be contaminated,

according to all applicable regulations.

Oil change procedure

+

Carry out the oil change only after the pump has

reached its normal operating temperature.

Switch off the pump and isolate it from its application.

Ventilate the pump to atmospheric pressure. Discon-

nect the electrical power cord.

+

Choose a suitable pad; oil may drip. Use a suitable

container to catch the oil.

Place a suitable container below the oil out-

let (9).

Remove the oil drain plug (9) below the rat-

Remove the oil drain plug (9) below the rat-

ing plate, paying attention to the O-ring.

Tilt the pump slightly and catch the oil in a

suitable container.

+

Dispose of the used oil according to regula-

Dispose of the used oil according to regula-

tions.

Screw in oil drain plug (9) with O-ring.

To flush the pump, pour in fresh oil (approx. 0.05 quarts

(50 ml)) through the pump inlet (1) (remove separator

if necessary).

Operate the pump briefly. Drain flushing oil and repeat

flushing procedure, if necessary, until all contamina-

tions are flushed out.

Remove the oil inlet plug (8). Fill in fresh oil through the

oil inlet port until the oil level reaches the mark ”max.”

(10). Do not overfill.

Screw in oil inlet plug (8).

Advertising