1. Disconnect the electrical power cord.

2. Adjust the

voltage selection switch to the supply voltage:

”115” corresponds to 100-120V and

”230” corresponds to 200-230V.

Voltage selection switch
+

Check that the voltage selection switch at the terminal box is

positioned correctly. Check every time before starting the pump.

Note: If the pump is switched on with wrong voltage selection,

the motor may be damaged!

Change the selection at the voltage selection switch only if the

pump is unplugged from the power source.

1

3

4

7

6

8

10

9

2

12

RE 2.5 / RZ 2.5 / RE 6 / RZ 6 with dual-voltage motor

(fig. RE 6)