Troubleshooting – VACUUBRAND RZ 9 User Manual

Page 39

Advertising
background image

page 39 of 49

Troubleshooting

Fault

Possible cause

Remedy

❑ Pump does not

start.

➨ Electrical power cord

not plugged in, electri-

cal supply failure?

✔ Plug in power cord.

Check fuse.

➨ Supply voltage does

not correspond with

the pump (see rating

plate)?

✔ Connect pump to a suit-

able supply voltage.

➨ Oil temperature below

54°F (12°C)?

✔ Operate the pump in suit-

able ambient conditions.

➨ Pump unit contami-

nated?

✔ Perform maintenance or

clean the pump unit.

➨ Motor thermally over-

loaded?

✔ Allow motor to cool down,

identify and eliminate

cause of failure. Manual

reset is necessary. Switch

off pump or unplug. Pro-

vide sufficient ventilation.

➨ Overpressure in outlet

line?

✔ Remove blockage in line,

open valve.

➨ Oil mist filter clogged

(filter element in con-

tact with oil)?

✔ Drain oil, replace filter

element if necessary.

❑ Pump does not

achieve its ultimate

vacuum or usual

pumping speed.

➨ Centring ring at small

flange connection not

correctly positioned, or

leak in the pipeline or

vacuum system?

✔ Check pump directly -

connect vacuum gauge

directly at pump inlet -

then check connection,

pipeline and vacuum

system if necessary.

➨ Long, narrow vacuum

line?

✔ Use lines with larger di-

ameter, length as short as

possible.

➨ Oil contaminated (also

with solvents)?

✔ Change oil and flush with

oil (see ”Oil change”).

➨ Not enough oil?

✔ Top up oil to correct level.

➨ Measuring procedure

or vacuum gauge not

suitable for the applica-

tion?

✔ Choose a suitable mea-

suring procedure or vacu-

um gauge.

➨ Pump has not yet

reached its operating

temperature?

✔ Allow the pump to reach

its operating temperature.