Replacing the device fuses – VACUUBRAND RZ 9 User Manual

Page 24

Advertising
background image

page 24 of 49

Replacing the device fuses

Unscrew the fuse holder at the terminal box

of the pump using a slotted screw driver.

Replace the defective fuse by a fuse of the

same type (see “Technical data”, pg. 17).

Reassemble holder with fuse to the pump.

Switch off the pump.


Disconnect the electrical power cord before un-

screwing the fuse holder.

Identify and eliminate the cause of failure before switch-

ing on the pump again.

• Double pole / neutral fusing:

The pump has two fuses and might have live circuits,

even when one fuse has blown or has been removed.

Switch off the pump or disconnect the electrical power

cord before replacing fuses.

5

13

Terminal box RE 2.5 / RZ 2.5 / RE 6 / RZ 6 with dual-voltage motor