Pulse outputs – Heath Consultants Heath Data Recorder (HDR) User Manual

Page 17

Advertising
background image

17

Pulse outputs

1. The HDR has two pulse outputs that can be independently configured for Alarm closure (for external visual or

audible alert or to notify another device of the alarm) or pulse repeating.
To set up the pulse outputs for alarm pulse closure:

a. Go to the Configure column.
b. Select “Configure user friendly names”.
c. Select the pulse output to activate, click “OK”.
d. Then go to the configure column/ “Configure Pulse I/O and 4-20 Ma Sender”.
e. In the “Pulse Outputs” area, click under “Use”, select “Alarm” to create a contact closure upon alarm.
f. Select “Pulse” to allow pulses to be sent (when input pulses are configured).

Note: The “Force On”= Contact closed. “Force Off”= Contact open. “Short On”= Can be used to control a motorized

regulator for remote pressure control.

2. To set up the pulse outputs for remote pressure control:

a. Go to the configure column/ “Configure Pulse I/O and 4-20 Ma Sender”.
b. In the “Pulse Outputs” area, click under “Use”, select “Short On”
c. In the “Momentary” area, enter the number of seconds to keep the pulse output closed (when the “Pulse

output button” is clicked on the Display 1 screen).

d. To set up the button on Display 1, go to “Configure column”/ “Configure User Friendly names”, click Pulse

Output 1 (for pulse 1) &/or Pulse Output 2 (for pulse 2). Enter a custom name if necessary (Example:
“Increase”/ “Decrease” or “Up”/ “Down”).

Note: After setting up pulse outputs for controlling a closure, Go to “Display 1” and click one of the four buttons at

the bottom of the screen. During the pulse closure, a scroll bar will appear displaying the progress of the
closure.



After clicking “Pulse output 1” a scroll bar appears

Configuring Pulse outputs for 5 second closure.

for 5 seconds during the “Closure”.