When calling the factory or your area representative for further troubleshooting, please have the following information
ready:

Customer Information

1. Name(s)

2. Company

3. Phone Number(s)

4. Location (site I.D.)

Hardware Information

1. Model of Instrument
2. Instrument Firmware Version
3. Battery Voltage

4.

Instrument Serial Number

5. Date of Manufacture
6. Transducer Serial Number(s) (If Applicable)
7. Any externally connected devices
8. Phone Number (If Applicable)
9. Most Recently Printed IOR Report (If Applicable)

Periodic Maintenance

Annual Checks

1. Verify calibration and recalibrate if required.
2. Verify proper battery main voltage (1.2 - 1.5 volts).
3. Verify proper battery Lithium backup voltage (3.4 - 3.6 volts).
4. Change batteries if necessary (Main batteries below.8V/ Lithium backup battery below 3.4V).
5. Check for leaks at pressure connections.

6.

Check for corrosion at wire terminations.

NOTE:

Heath Consultants Incorporated recommends annual checks on these instruments.