Controller set-up – Labconco SteamScrubbe & FlaskScrubbe Glassware Washers 45781 Series User Manual

Page 22

Advertising
background image

Chapter 3: Getting Started

Product Service 1-800-522-7658

18

18

4.

Remove the cover over the electrical junction box by removing the two

screws.


5.

Make sure electrical power is turned off before making any connections.

6.

On 115V models, connect the customer-supplied power wires to the black and

white wires, using wire connectors. On 230V models, attach customer-
supplied power wires to the brown and blue wires, using wire connectors.
Customer-supplied wires must be clamped to the hole where they pass
through the rear of the junction box, if prescribed by local codes. If it is
necessary to use a bushing with an integral clamp at the inlet to the junction
box, remove the plastic bushing supplied with the washer before installing the
new customer supplied bushing.

7.

Connect the customer-supplied ground wire to the green screw.

8.

Replace the cover on the electrical junction box.

9.

Replace the lower panel and door panel assembly on the washer.

Controller Set-Up

The washer control system has been set at the factory to agree with the
configuration of the washer as it was shipped. The control may be customized or
modified if certain optional accessories are added.

Advertising