Chapter 6: Modifying Your Washer

Product Service 1-800-522-7658

41

Installing Small Spindles

Ten small spindles are included with each Spindle Rack. Since the large volume
of water which flows through the standard spindles may be great enough to force
the glassware off the spindle, small spindles should be used when small pieces of
glassware will be washed.

To install small spindles:

1.

Remove the standard spindle by inserting the tool supplied with the Spindle

Rack or a screwdriver into the slot on the top and turning counterclockwise.

2.

Replace the standard spindle with a small spindle.

Installing the Optional Cool Drain Water Kit

An optional Cool Drain Water Kit is available to reduce the water temperature of
the drain water in installations where hot drain water is not permitted. It may be
installed on Undercounter or Freestanding models. A supply of cold water must
be available. It is recommended that the cold water temperature is colder than
54°F (12°C). Warmer water temperature will cause the drain temperature to be
higher. The cold water supply must be capable of a flow rate of 3.5 GPM (13.2
LPM) and a pressure of 20-120 PSI (138-827 kPa). Hot drain water will be mixed
with cold water during drain operations. To install the optional Cool Drain Water
Kit, refer to the figure below and the following instructions.

To install the optional Cool Drain Water Kit:

1.

Disconnect power to the washer.

2.

Remove the back panel of the washer if it is a Freestanding model or the cover

panel of an Undercounter model.

3.

Attach the water valve assembly to the rear flange of the base using the screws

provided.

4.

Connect the wires coming from the water valve assembly to the mating

connector in the wiring harness of the washer.

5.

Connect the cold water supply to the water valve assembly. Do not allow the

tube to kink when the washer is placed into its operational location. If
necessary, replace the straight fitting on the cool water valve with the supplied
elbow.

6.

Replace the back panel if it is a Freestanding model or the cover panel if it is

an Undercounter model.

7.

Attach the hose from the valve to a Twin Inlet High Flow Air-gap suitable for

dishwasher applications. This may be obtained from a plumbing supply store.

8.

Attach the drain hose from the Glassware Washer to the other inlet of the Air-

gap.

9.

Attach the Air-gap to the building drain.

10.

Turn on the electrical power.

Advertising