Chapter 7: Troubleshooting

Product Service 1-800-522-7658

49


PROBLEM

CAUSE

CORRECTIVE ACTION

Defective hose to fill
switch

Check for splits.

Check for loose clamps.

Defective fill level
switch

Recalibrate switch. Contact Labconco Product
Service.

Replace switch.

Defective controller

Check output from PCB to both valves and DI
pump.

Replace PCB. Contact Labconco Product
Service.

Holes in filter screen
too small

Contact Labconco Product Service.

Defective grommet
where hose connects
to fill level switch

Replace grommet.

Low Temp Alarm

Heater inoperative

Check electrical connections.

Check voltage to heater.

Check resistance of heater. If open circuit,
replace heater.

Defective controller

Check output from PCB to heater.

Replace PCB. Contact Labconco Product Service.

Defective
temperature probe

Check electrical connections.

Contact Labconco Product Service. Replace if
necessary.

Defective high
temperature cut out
switch

Check resistance of safety cut out switch. If open
circuit at room temperature, replace switch.