Do not permit any

uncontrolled pressurizing. Make

sure that the exhaust pipeline cannot become blocked.

If there is an exhaust isolation valve, make sure that you

cannot operate the equipment with the valve closed to

avoid a

risk of bursting!

+

Always provide a free and pressureless exhaust outlet

to avoid damage to pump valves and risk of bursting of

condenser glassware.

+

Check the overpressure safety relief device at the ex-

haust waste vapor condenser at appropriate intervals.

+

Keep the electrical power cord away from heated sur-

faces.

• Provide a firm, level platform for the equipment. Check

that the system which you are going to evacuate is

mechanically stable. Check that all fittings are secure.

Ensure a stable position of the pump without any me-

chanical contact other than the pump feet.

• Comply with

maximum permissible pressures at in-

let and outlet and with

maximum permissible pres-

sure differences between inlet and outlet. See section

”Technical data”, pg. 24. Do not operate the pump

with overpressure at the inlet.

• Avoid overpressure of more than 17.5 psi absolute (1.2

bar absolute) in the event that gas or inert gas is connect-

ed to the pump, to the gas ballast or to a venting valve.

Note: Flexible elements will shrink when evacuated.

• Connect hoses gas tight at inlet and outlet of the pump.

• Ensure that no foreign objects can be drawn into the

pump.

• Check the power source and the pump’s rating plate

to be sure that the power source and the equipment

match in voltage, phase, and frequency.

• On pumps with a dual-voltage motor, check that the

voltage selection switch is set correctly. Do not change

the setting of the voltage selection switch while the