Technical data, Echnical data”, pg. 24) – Labconco VACUUBRAND Diaphragm Pump 7393000 User Manual

Page 24

Advertising
background image

page 24 of 74

Technical data

General technical data valid for all pumps/pumping units

Maximum permissible inlet pressure

(absolute)

psi

(bar)

16

(1.1)

Maximum permissible outlet pressure

(absolute)

psi

(bar)

16

(1.1)

Maximum pressure difference between

inlet and outlet

psi

(bar)

16

(1.1)

Maximum permissible pressure (abso-

lute) at gas ballast valve

psi

(bar)

17.5

(1.2)

Permissible ambient temperature

storage / operation

°F

(°C)

14 to 140 / 50 to 104

(-10 to +60 / +10 to +40)

Permissible relative atmospheric mois-

ture during operation (no condensation)

%

30 to 85

No-load speed 50/60 Hz

rpm

1500 / 1800

Device fuse

slow blow fuse 6.3 A

Motor protection

thermal cutout, manual reset

Degree of protection IEC 529

IP 40

Coolant connection

(waste vapor condenser, only ”EK”)

hose nozzle for tubing

I.D. 1/4” - 5/16”

(hose nozzle DN 6-8 mm)

Maximum permissible pressure of

coolant at waste vapor condenser (”EK”)

psi

(bar)

87 (absolute)

(6 (absolute))

Permissible range of coolant temperature

(waste vapor condenser, only ”EK”)

°F

(°C)

5 to 68

(-15 to +20)

Volume of catchpot (only ”AK” / ”EK”)

quarts

(ml)

0.52

(500)

A-weighted emission sound pressure

level* (uncertainty K

pA

: 3 dB(A))

dB(A)

45

* Measurement according to EN ISO 2151:2004 and EN ISO 3744:1995 at 230V/50Hz and ulti-

mate vacuum with exhaust tube at outlet.

Advertising