Position clamping bracket (S) with counter-

sunk bores facing upwards.

Align the countersunk bores with the thread-

ed pegs.

Loosely fasten the countersunk screws and

correct the alignment of the valve heads if

necessary.

Tighten countersunk screws with Torx screwdriver T20.

+

Torque: 2.2 ft

.

lb

f

(3 Nm).

Bring the diaphragms (F) into a position, in

which they are in contact with the housing

(C) and centered with respect to the bore.

Put on head cover (J) with valve heads (O)

and connections attached.

+

Pay attention to the correct orientation of the

head covers:

Housing with head alignment pin: The head

alignment pin (A) at the pump housing (C)

has to fit into the recess at the head cover

(J).

Housing with mark (A): Align the recess at

the head cover with the mark at the pump

housing.

Loosely screw in the Allen head screws (H)

at the head covers diagonally at first slightly

with a 5 mm wide Allen key, then tighten.

+

Recommended torque: 8.9 ft

.

lb

f

(12 Nm).

Slide the caps (I) into the head cover.

Valve head with gas ballast or hose nozzle connection:

Insert square nut (K) in the groove of the head cover (J) or position

square nut in the groove and then screw on connection fastener.

+

Loosely fasten fillister head screw (L).