Installing a pump in a vacuum system

If dangerous or polluting fluids could be released at the

outlet, install an appropriate system to catch and dis-

pose of those fluids.

+

Connect a gas-tight exhaust line at the pump outlet if

necessary. Always vent exhaust gases appropriately

(e.g., into a fume hood).

+

Never block the gas outlet. The exhaust line must al-

ways be free of obstructions (no back pressure) to en-

sure an unimpeded discharge of gas. The cross-sec-

tion of the outlet tubing must be at least the size of the

pump’s exhaust connection.

+

Particles and dust must not be aspirated. If necessary,

you must install appropriate filters. You must ensure

their suitability concerning gas flow, chemical resis-

tance and resistance to clogging prior to use.

+

Make sure ventilation is adequate, especially if the pump

is installed in an enclosure, or if the ambient temperature

is elevated. Provide external ventilation, if necessary.

• Reduce the transmission of vibration. Prevent mechan-

ical load due to rigid pipelines. Insert elastic hoses or

flexible elements as couplings between the pump and

rigid pipes.

Note: Flexible elements will compress or

flatten when evacuated if not designed for use under

vacuum.

• Hose connections at the pump inlet must always be

gas tight.

• A power failure may cause accidental ventilation of the

pump, especially if the gas ballast valve is open . If this

constitutes a potential source of danger, take appropri-

ate safety measures.