Replacing the overpressure safety relief device – Labconco VACUUBRAND Diaphragm Pump 7393000 User Manual

Page 65

Advertising
background image

page 65 of 74

Remove the 4 screws affixing the counter

holder of the exhaust vapor condenser with

a Torx driver T10. Remove the exhaust va-

por condenser. In doing so pull the PTFE

hose out of the inlet of the exhaust vapor

condenser.

If necessary, pull off the old pressure relief

valve and install the new one. Check for cor-

rect position of the PTFE-foil under the pres-

sure relief valve.

Feed the PTFE hose into the inlet of the ex-

haust vapor condenser. Mount the condens-

er with the counter holder (Torx screws T10)

to the pump. Fasten union nut.

Overpressure

safety relief device ............................ 638821

(at exhaust waste vapor condenser)

Remove catchpot (9) at outlet (see ”Use and

operation”, pg. 38).

Loosen the union nut at the inlet of the ex-

haust vapor condenser.

Replacing the overpressure safety relief device