Notice – Labconco VACUUBRAND Diaphragm Pump 7393000 User Manual

Page 43

Advertising
background image

page 43 of 74

its. Clean the pump if necessary to avoid an increase of

the pump’s operating temperature.

Operation with silencer (optional) at the outlet: Operating

the pump at a high inlet pressure or pumping dusty gases

for a long time may cause clogging of the silencer. Check

the silencer regularly and replace if necessary.

In case of overload, the motor is shut down by a

self-hold

thermal circuit breaker in the winding.

Note: Only a manual reset is possible. Switch off the

pump and disconnect the electrical power cord. Identify

and eliminate the cause of failure. Wait approximately five

minutes before restarting the pump.

Note: In case of supply voltage below 100V, the lock

of the breaker may not latch and the pump might restart

on its own after sufficient cooling. Take appropriate pre-

cautions, if an automatic restart of the pump may lead

to a dangerous situation.

A warm up period (approximately 15 min.) is required to

ensure that the rated ultimate vacuum and pumping speed

are attained. Avoid overheating (e.g., due to hot process

gases).

Pumps with flow control diaphragm valve:

Use the flow control diaphragm valve (14) at the pump

inlet to control the pumping speed. Open flow control

diaphragm to pump down.

Opening the flow control diaphragm: Turn counterclock-

wise. Do not attempt to open the valve further than when

resistance is first encountered.

Closing: Turn clockwise. Close flow control diaphragm

valve slightly hand-tight. Further tightening will not in-

crease sealing of the valve, and may lead to damage.

Note: Over-tightening or -loosening may damage the dia-

phragm or the valve seat, and the valve may not close

properly thereafter.

Replace diaphragm in case of leaks.

NOTICE

NOTICE

Advertising
This manual is related to the following products: