To reduce pump noise emanating from the pump exhaust

port, connect an exhaust hose or use a silencer (see ”Ac-

cessories”, pg. 48).

Separator (inlet) and

exhaust waste vapor condenser (outlet)

Assembling the hose nozzle with union nut:

Take the hose nozzle with attached compression fer-

rule and union nut out of the catchpot and put onto inlet

connection (version 2 AK: onto inlet and outlet connec-

tions).

Tighten the union nut by hand until you can feel the

stop. Then tighten an additional 1/4 rotation with an

open-ended wrench (size 17mm) for final installation.

Catchpots:

The catchpot at the inlet protects against drop-

lets and particles from entering the pump.
+

Enhances lifetimes of diaphragms and

valves.

+

Improves vacuum performance in applica-

tions with condensable vapors.

Both catchpots are coated with a protective lay-

er to protect against shattering in case of break-

age or implosion.

Assemble the catchpots at the inlet and at

the outlet using joint clips.

Exhaust waste vapor condenser:

Assemble the hose nozzles for coolant inlet

(12) and coolant outlet (13) tubing at the ex-

haust waste vapor condenser.

The

exhaust waste vapor condenser enables

an efficient condensation of the pumped vapors

at the outlet.
+

No backflow of condensates.

+

Controlled recovery of condensates.

+

Close to 100% solvent recovery.

+

The isolation cover protects against glass

overpressure

safety relief

device (11)

catchpot

at outlet (9)

catchpot

at inlet (9)

outlet (gas!) (2)

(12)

(13)

Advertising