Gas inlet temperatures – Labconco VACUUBRAND Diaphragm Pump 7393000 User Manual

Page 25

Advertising
background image

page 25 of 74

Gas inlet temperatures

Type

ME 2C NT

ME 4C NT

ME 4C NT + 2 AK

Maximum pumping

speed* 50/60 Hz

(ISO 21360)

cfm

(m

3

/h)

1.2 / 1.4

(2.1 / 2.4)

2.3 / 2.6

(3.9 / 4.3)

Ultimate vacuum

(absolute)

Torr

(mbar)

52.5

(70)

52.5

(70)

Rated motor power

hp (kW)

0.24 (0.18)

Maximum permissible range

of supply voltage ( ±10% )

Attention: Observe specifications

of rating plate!

100-115 V~ 50/60 Hz, 120V~ 60 Hz

230 V~ 50/60 Hz

Dual voltage motor

100-115 V~ 50/60 Hz, 120~ V 60 Hz /

200-230 V~ 50/60 Hz

Maximum rated current at:

100-115 V~ 50/60 Hz, A

120 V~ 60 Hz

200-230 V~ 50/60 Hz A

230 V~ 50/60 Hz A

3.4

1.8

1.8

Inlet

hose nozzle for tubing I.D. 3/8" or small flange

KF 16 (hose nozzle DN 10 mm or small flange

KF 16)

Outlet

hose nozzle for tubing I.D. 3/8”

(hose nozzle DN 10 mm)

Dimensions L x W x H

approx.

design 2 AK

in

(mm)

in

(mm)

9.6 x 8.3 x 7.8

(243 x 211 x 198

-

(-)

10.0 x 9.6 x 7.8

(254 x 243 x 198)

12.4 x 9.6 x 11.5

(316 x 243 x 291)

Weight approx.

design 2 AK

lbs. (kg)

lbs. (kg)

22.5 (10.2)

-

24.3 (11.1)

30.0 (13.6)

* Pumping speed of diaphragm pump

Operating condition

Inlet pressure

Permitted range of gas

temperatures at inlet

Continuous operation

> 75 Torr (100 mbar)

(high gas load)

50 °F to 104 °F

(+10°C to +40°C)

Continuous operation

< 75 Torr (100 mbar)

(low gas load)

32 °F to 140 °F*

(0°C to +60°C*)

Short-time

(< 5 minutes)

< 75 Torr (100 mbar)

(low gas load)

14 °F to 176 °F*

(-10°C to +80°C*)

* if pumping potentially explosive atmospheres: 50 °F to 104 °F (+10°C to +40°C)