During operation

Vent and dispose of

potentially dangerous gases or

vapors at the outlet of the pump appropriately.

+

Due to the high compression ratio, the pump might gen-

erate overpressure at the outlet. Check pressure com-

patibility with system components (e.g., exhaust tubing

or exhaust valve) at the outlet. Ensure that the pump

outlet is neither blocked nor restricted.

+

Maximum ambient temperature: 104 °F (40 °C)

Check the maximum temperatures, if installing the

pump in a cabinet or a housing. Make sure ventilation

is adequate, especially if the ambient temperature is

elevated.

• If the pump is installed at an altitude of more than 3300

ft (1000 m) above mean sea level, check compatibil-

ity with applicable safety requirements, especially IEC

60034. There is a risk of the motor overheating due to

insufficient cooling.

• Check compatibility with the

maximally permitted

pressure at outlet and the maximum pressure differ-

ence between inlet and outlet ports.

Do not start the pump if the

pressure difference between

inlet and outlet ports exceeds max. 16.0 psi (1.1 bar).

Attempts to start the pump at higher pressure difference

may cause stalling and damage of the motor.

If pumping condensable vapors (water vapor, solvents,

etc.), let the pump run with

gas ballast to help purge any

condensation in the pump.

Prevent internal condensation, transfer of liquids or dust.

The diaphragms and valves will be damaged, if liquids are

pumped in significant amounts.

Check the pump regularly for external soiling and depos-