Position new diaphragm (F) between dia-

phragm clamping disc with square head

screw (G) and diaphragm support disc (E).

+

Note: Position diaphragm with pale side to-

wards diaphragm clamping disc (facing pump

chamber).

+

Make sure that the square head screw of the diaphragm clamping disc

is correctly seated in the guide hole of the diaphragm support disc.

Lift the diaphragm at the side. Place the diaphragm carefully together

with diaphragm clamping disc and diaphragm support disc in the dia-

phragm key.

+

Avoid damage of the diaphragm: Do not excessively bend or crease

the diaphragm.

Check for washers (D) between the diaphragm support disc (E) and

the connecting rod (B). Do not mix the washers from the different pump

heads, since these are set at the factory to ensure proper pump per-

formance. Make sure that the original number is reassembled at the

individual pump head.

+

Too few washers: The pump will not attain vacuum specification. Too

many washers: Diaphragm clamping disc will hit head cover, causing

noisy operation and possibly causing the pump to seize up.

+

If the old diaphragm is difficult to separate from the diaphragm support

disc, immerse assembly in naphtha or petroleum ether. Do not inhale

vapors!

+

Check diaphragm (F) for damage and re-

place if necessary.

Lift diaphragm carefully sidewise.

+

Never use a pointed or sharp-edged tool to

lift the diaphragm.

Use the diaphragm key to grip the diaphragm

support disc (E) below the diaphragm.

Unscrew diaphragm support disc (E) with diaphragm (F) and dia-

phragm clamping disc (G).

F

G

G

E

F