Operating conditions – Labconco VACUUBRAND HYBRID Rotary Vane Pump 7584000 User Manual

Page 11

Advertising
background image

page 11 of 56

Operating conditions

These pumps are not approved for operation in po-

tentially explosive atmospheres.

Do not operate the

pumps in potentially explosive atmospheres.

These pumps are not approved for the pumping of po-

tentially explosive atmospheres.

Do not pump poten-

tially explosive atmospheres.

The pumps are

not suitable to pump any of the sub-

stances listed below.

Do not pump:

-

unstable substances

- substances which react explosively under

impact

(mechanical stress) without air

- substances which react explosively when being ex-

posed to

elevated temperatures without air,

-

self inflammable substances,

- substances which are

inflammable without air

-

explosive substances.

The pumps are

not approved for operation below

ground.

Do not operate the pump below ground.

+

The pumps are

not suitable for pumping dust.

Do not pump dust.

Do not pump substances which may form deposits in-

side the pump. The pumps are not suitable for pumping

substances which may form deposits inside the pump.

Deposits and condensate in the pump may lead to in-

creased temperatures even to the point of exceeding

the maximum permitted temperatures. Deposits may

cause seizing of the pump unit.

• Check the inlet and outlet of the pump and its oil condi-

tion, if there is a danger of forming

deposits inside the

pump, e.g., in the pump unit of the rotary vane pump or

in the pump chamber of the diaphragm pump. Inspect

the pump regularly and clean if necessary.