page 9 of 56

• Check the power source and the pump’s rating plate

to be sure that the power source and the equipment

match in voltage, phase, and frequency.

• Use only oil of the recommended type. Other oils or op-

erating fluids may cause damage of the pump or dan-

ger.

Use

special oils (see “Oil change”, pg. 32) for the

rotary vane pump, if, e.g.,

- acid vapors are pumped.

- extremely high aging stability is required.

- oxygen or other strong oxidizing agents are pumped.

• Take adequate precautions when handling pump fluids,

lubricants, and solvents. Use appropriate protective

clothing, safety goggles and protective gloves to avoid

excessive contact with the skin and possible skin irrita-

tions (including dermatitis). Do not inhale or swallow!

Ensure that the pump location is well ventilated and

that possible toxic effects of certain vapors are avoided.

Maintain adequate levels of hygiene and cleanliness.

• Comply with all relevant statutory requirements and

regulations concerning the handling, storage and dis-

posal of oil.

Note: Do not allow oils to enter the drainage system

or other bodies of water. Spillage can cause accidents

(slip hazard)!

Keep a minimum distance of 8 in (20 cm) between the

cooling fan and surrounding items (e.g., housing, walls,

etc.). Check fan regularly for dust/dirt. Clean fan guard if

necessary to avoid a reduction of ventilation.

The On/Off switch is located at the side of the terminal

box.

Use only hoses at the inlet and outlet of the pump with

an inner diameter at least as large as the diameter of the

pump’s connections (to avoid overpressure at the outlet,

NOTICE