Fault

Possible cause

Remedy

❑ Oil leakage.

➨ Oil spilled?

✔ Absorb the oil and dis-

pose of according to reg-

ulations.

➨ Other cause?

✔ Contact local distributor.

❑ High oil consump-

tion.

➨ High inlet pressure?

✔ Normal. Top off oil when-

ever necessary. Install oil

mist filter at outlet.

➨ Operation with gas bal-

last?

✔ Top off oil whenever nec-

essary. Install oil mist filter

at outlet.

➨ Oil level too high?

✔ Lower oil level to mark

”max.”.

❑ Fast aging of the

oil.

➨ Pumping aggressive

gases?

✔ Use suitable oil.

➨ Condensation in the

pump?

✔ Use separator or cold

trap.

➨ Oil mist filter defective? ✔ Replace filter. Check

safety pop valve.

➨ Diaphragm pump does

not reach its pumping

speed?

✔ Perform maintenance of

diaphragm pump.

❑ Pump seized.

➨ Ambient temperature

too high ( > 104 °F

(40°C))?

✔ Provide sufficient ventila-

tion.

➨ None of the above

mentioned causes?

✔ Contact local distributor.