Gas inlet temperatures, Type rc 6 – Labconco VACUUBRAND HYBRID Rotary Vane Pump 7584000 User Manual

Page 19

Advertising
background image

page 19 of 56

Type

RC 6

Rated current during operation

f.)

at:

100-120 V~ 50/60 Hz

230 V~ 50/60 Hz

A

A

5.7 / 5.1

2.7 / 2.6

Degree of protection IEC 529

IP 40

A-weighted emission sound

pressure level

g.)

(uncertainty K

pA

: 3 dB(A))

dB(A)

50

Inlet

small flange KF DN 16

Outlet

hose nozzle for tubing I.D. 3/8”

(hose nozzle DN 10 mm)

Dimensions L x W x H approx.

in

(mm)

20.1 x 12.0 x 9.1

(510 x 305 x 230)

Weight with oil filling approx.

lbs.

(kg)

53.4

(24.2)

a.) Partial pressure of permanent gases measured at pump inlet.

b.) The total pressure is higher than the partial pressure because the former includes the vapor

pressure of the pump oil and other condensable vapors (e.g., water); the condition of the oil

(cleanliness, content of hydrocarbons with higher partial pressures) is crucial for this value.

c.) The maximum inlet pressure for water vapor, or rather the maximum inlet pressure for vapor,

cannot be specified in accordance with ISO 21360-2 because it cannot be determined for the

HYBRID pump in accordance with this standard. Because the diaphragm pump reduces the

pressure in the oil-sealed part of the RC 6, however, the water vapor tolerance is considerably

higher than that of a conventional oil-sealed rotary-vane pump.

d.) With inlet sealed and without gas ballast.

e.) Oil temperature and pressure in the oil reservoir are the relevant parameters for maximum

vapor inlet pressure and chemical resistance.

f.) During the start-up period of the pump (first 6 minutes after switching-on), the current draw

may be elevated to twice the nominal current draw during normal running conditions.

g.) Measurement according to EN ISO 2151:2004 and EN ISO 3744:1995 at 230V/50Hz and

ultimate vacuum with exhaust tube at outlet.

Gas inlet temperatures

Operating condition

Inlet pressure

Permitted range of gas

temperatures at inlet

Continuous operation

> 75 Torr (100 mbar)

(high gas load)

50 °F to 104 °F

(+10°C to +40°C)

Continuous operation

< 75 Torr (100 mbar)

(low gas load)

32 °F to 140 °F

(0°C to +60°C)

Short-time

(< 5 minutes)

< 75 Torr (100 mbar)

(low gas load)

14 °F to 176 °F

(-10°C to +80°C)

We reserve the right for technical modification without prior notice!