Technical data

Type

RC 6

Maximum pumping speed

50/60 Hz (ISO 21360)

cfm

(m

3

/h)

3.5 / 4.1

(5.9 / 6.9)

Ultimate partial pressure

without gas ballast

a.)

Torr

(mbar)

3*10

-4

(4*10

-4

)

Ultimate total pressure

without gas ballast

b.)

Torr

(mbar)

1.5*10

-3

(2*10

-3

)

Ultimate total pressure

with gas ballast

Torr

(mbar)

0.75*10

-2

(1*10

-2

)

Water vapor tolerance

c.)

Torr

(mbar)

>>30

(>>40)

Oil capacity min./max.

quarts

(ml)

0.36 / 0.56

(340 / 530)

Total pressure in oil reservoir

d.)

Torr

(mbar)

13.5

(18)

Oil temperature

e.)

(under typical

operating conditions)

°F

(°C)

140

(60)

Recommended oil

VACUUBRAND B-oil

Maximum permissible inlet / outlet

pressure (absolute)

psi

(bar)

16

(1.1)

Maximum permissible pressure

difference between inlet and outlet

psi

(bar)

16

(1.1)

Maximum permissible pressure

(absolute) at gas ballast valve

psi

(bar)

17.5

(1.2)

Permissible ambient temperature

storage / operation

°F

(°C)

14 to 140 / 54 to 104

(-10 to +60 / +12 to +40)

Permissible relative atmospheric

moisture during operation

(no condensation)

%

30 to 85

Rated motor power

hp

(kW)

0.5

(0.37)

No-load speed 50/60 Hz

rpm

1500 / 1800

Maximum permissible range of

supply voltage

Attention: Observe specifications

of rating plate!

100-120 V~ +5%/-10% 50/60 Hz

230 V~

±10% 50/60 Hz

Motor protection

self-hold thermal cutout,

manual reset

Advertising