Troubleshooting – Labconco VACUUBRAND HYBRID Rotary Vane Pump 7584000 User Manual

Page 38

Advertising
background image

page 38 of 56

Troubleshooting

Fault

Possible cause

Remedy

❑ Pump does not

start.

➨ Electrical power cord

not plugged in, electri-

cal supply failure?

✔ Plug in power cord.

Check fuse.

➨ Supply voltage does

not correspond with

the pump (see rating

plate)?

✔ Connect pump to a suit-

able supply voltage.

➨ Start-up control defec-

tive?

✔ Contact local distributor.

➨ Oil temperature below

54°F (12°C)?

✔ Operate the pump in suit-

able ambient conditions.

➨ Pump unit of rotary

vane pump contami-

nated?

✔ Perform maintenance or

clean the pump unit.

➨ Motor thermally over-

loaded?

✔ Allow motor to cool down

(< 50°C (122°F)), iden-

tify and eliminate cause

of failure. Manual reset

is necessary. Switch off

pump or unplug. Provide

sufficient ventilation.

➨ Overpressure in outlet

line?

✔ Remove blockage in line,

open valve.

➨ Oil mist filter clogged

(filter element in con-

tact with oil) or over-

filled?

✔ Drain oil, replace filter

element if necessary.

➨ Pump seized?

✔ See below.

❑ Pump does not

achieve its ultimate

vacuum or usual

pumping speed.

➨ Measuring procedure

or vacuum gauge not

suitable for the applica-

tion?

✔ Choose a suitable mea-

suring procedure or vacu-

um gauge.

➨ Inlet blocked?

✔ Open inlet. Ensure that

the inlet line does not be-

come blocked.

➨ Centring ring at small

flange connection not

correctly positioned?

✔ Check small flange con-

nections.

➨ Long, narrow vacuum

line?

✔ Use lines with larger di-

ameter, length as short as

possible.