Cleaning and inspecting the diaphragm pump heads – Labconco VACUUBRAND HYBRID Rotary Vane Pump 7584000 User Manual

Page 43

Advertising
background image

page 43 of 56

Drain the catch pot of the oil mist filter before beginning maintenance

to prevent leaking of condensate!

Drain oil (see ” Oil change”) to prevent oil from making its way into the

filter element. Refill pump with oil before restarting the pump!

For maintenance, lay the pump on its side with the pump head to be

maintained at the top. Support the pump appropriately.

Attention: Prevent any damage to the oil mist filter or to the manome-

ter. Do not use the oil mist filter to prop up the pump.

Loosen the union nuts of the hose connec-

tions at the pump head with an open-ended

wrench (17 mm).

Tools required (metric):
- Phillips screw driver size 2

- Open-ended wrench 14/17 mm

- 5 mm wide Allen key

- Diaphragm key width 66 mm

Diaphragm ....................................................................................................639786

Valve .............................................................................................................638440

Vacuum gauge DVR 2..................................................................................682902

Diaphragm key (width 66 mm) ....................................................................636554

Cleaning and inspecting the diaphragm pump heads

Remove the cover plate. Use a Phillips screw

driver size 2 to loosen both screws (A). Pay

attention to lock washers.

A