During operation – Labconco VACUUBRAND HYBRID Rotary Vane Pump 7584000 User Manual

Page 25

Advertising
background image

page 25 of 56

with hose clamps, to protect against accidental detach-

ment.

Note: The pump must only be operated in a horizontal

position. When transporting a pump containing oil, take

great care not to tilt the pump to an angle that would allow

oil to make its way into the filter element.

During operation

Vent and dispose of

potentially dangerous gases or

vapors at the outlet of the pump appropriately.

+

Due to the high compression ratio, the pump might gen-

erate overpressure at the outlet. Check pressure com-

patibility with system components (e.g., exhaust tubing

or exhaust valve) at the outlet. Ensure that the pump

outlet is neither blocked nor restricted.

+

Maximum ambient temperature: 104 °F (40 °C)

Check the maximum temperatures, if installing the

pump in a cabinet or a housing. Make sure ventilation

is adequate, especially if the ambient temperature is

elevated.

• If the pump is installed at an altitude of more than 6500

ft (2000 m) above mean sea level, check compatibility

with applicable safety requirements, and adopt suitable

measures. There is a risk of the motor overheating due

to insufficient cooling.

• Check compatibility with the

maximally permitted

pressures at inlet and outlet and the maximum pres-

sure difference between inlet and outlet ports.

• Check the

oil level of the rotary vane pump every time

before starting the pump, however at least once a week.

Check the oil level more frequently if high amounts of

gas or vapor are pumped.