page 48 of 56

Assemble the cover plate. Pay attention to lock washers. Put the pump

in normal operating position.

+

Attention: Make sure that the pump is filled with oil before switching it

on, refill oil if necessary (see ”Oil change”, pg. 32)!

Check the operability of the diaphragm pump by measuring the pressure

in the oil reservoir:

Unscrew the manometer with its adapter (open end wrench width 17

mm). Remove the hose nozzle from the distributor plate (outlet) and

install at the oil inlet. Make sure the O-ring is seated correctly.

Using a suitable vacuum gauge (e.g., DVR 2) and making sure that it

is correctly calibrated, measure the pressure in the oil reservoir. When

the pump operates with the inlet line closed, the measured pressure in

the oil reservoir should be less than 25 mbar (19 Torr) (with gas ballast

valve of rotary vane pump closed).

Remove the hose nozzle from oil inlet. Reinstall it in the distributor

plate and reassemble the manometer with O-ring.

Assembling the connecting hose

Slip the hoses onto the hose connectors by

turning the fittings with an open-ended 14

mm wrench.

Tighten the union nuts of the hose connec-

tions at the pump heads with an open-ended

17 mm wrench.

+

Tighten union nuts first by hand and then

tighten one full turn using the open ended

wrench.

Replace diaphragm and valves of the opposite side of the

pump in the same way.

Checking the operability of the diaphragm pump