Shutdown, Notice – Labconco VACUUBRAND HYBRID Rotary Vane Pump 7584000 User Manual

Page 31

Advertising
background image

page 31 of 56

open and then with the gas ballast valve closed, after the

pump has been isolated from the vacuum application.

Prevent internal condensation, transfer of liquids or dust.

The pump requires regular inspection and maintenance.

Shutdown

Has the pump been exposed to condensate?

- Allow the pump to continue to run for a few minutes

with the rotary vane pump’s gas ballast valve open be-

fore switching off.

Long-term:

- Separate the pump from the application.

- Drain condensates (catchpots of the separator at the

inlet and of the oil mist filter).

- Flush the pump with dry nitrogen.

- Carry out oil change (see “Oil change”, pg. 32).

- Fill the pump completely with new oil (to above the

”max.” mark!).

Attention: Before restarting the pump, drain oil to the

maximum oil level (”max.” mark)!

- Close the manual gas ballast valve.

- Close inlet and outlet ports (e.g., with transport caps or

blind flanges).

- Store the pump under dry conditions.

- Carry out oil change and maintenance prior to use if the

pump has been stored for longer than one year.

Note: When transporting a pump containing oil, take great

care not to tilt the pump to an angle that would allow oil to

make its way into the filter element. Drain oil if necessary.