Under normal operating conditions:

- Check oil level every time before starting the pump.

- Carry out an oil change once aging of the oil is indi-

cated by darker color compared to new oil.

- Carry out oil changes at least once a year.

- Dispose of the used oil, which may be contaminated,

according to all applicable regulations.

If the oil contains small quantities of water/solvent:
+

The oil can be cleaned to a certain extent by operating

the pump for 1 - 2 hours with the inlet line sealed and

the gas ballast valve open.

Oil change procedure

+

Carry out the oil change only after the pump has reached

its normal operating temperature.

Switch off the pump and isolate it from its application.

Ventilate the pump to atmospheric pressure. Discon-

nect the electrical power cord. To vent the oil reservoir,

remove the manometer with its adapter carefully (open

end wrench width 17 mm). Pay attention to the O-ring!

Do not turn the manometer itself as leaks might be

caused!

+

Choose a suitable pad; oil may drip. Use a suitable

container to catch the oil.

Place a suitable container below the oil outlet

(9).

Remove the oil drain plug (9) below the rat-

ing plate, paying attention to the O-ring.

Tilt the pump slightly and catch the oil in a

suitable container.

+

Dispose of the used oil according to regula-

tions.

Screw in oil drain plug (9) with O-ring.

Screw in manometer with O-ring.

To flush the pump, pour in fresh oil (approx. 0.05 quarts

(50 ml)) through the pump inlet (1) (remove separator if

necessary).