3 Safety

20

Version 04/2012, Rev. 1.4 of 25/11/2013 • sb

Translation of the original operating manual

Pos: 47 / 200 Christ /37 0 RVC-BA (STANDARDM ODULE)/ 030 Sich erh eit/0 30- 006 0-0 030 Bra ndsch utz @ 7\m od_ 130 933 104 011 5_68 .docx @ 4 646 3 @ 3 @ 1

3.6.3 Fire prevention

Fuses protect certain electrical circuits within the RVC against over-current
conditions.

GEFAHR

• Always use fuses of the same type and rating.
• Do not evaporate explosive or inflammable substances.
• Do not use the RVC within hazardous locations where there is a risk of

explosion.

Pos: 48 / 010 Unive rsalm odul e/ L ee rzeile @ 0\ mod _120 211 624 450 0_0. docx @ 11 4 @ @ 1

Pos: 49 / 200 Christ /37 0 RVC-BA (STANDARDM ODULE)/ 030 Sich erh eit/0 30- 006 0-0 040 T her misch e Siche rheit @ 7\ mod _13 093 310 4047 4_6 8.d ocx @ 464 71 @ 3 @ 1

3.6.4 Thermal safety

During the evaporation process, the housing, lid and interior of the RVC
can reach surface temperatures of more than +50°C.

DANGER

• Open and close the lid at the heat-insulated handle!
• Ensure that no limbs come into contact with hot unit parts or

accessories. Danger of burns!

• Wear heat-resistant gloves when inserting or removing the rotors.

Pos: 50 / 010 Unive rsalm odul e/ L ee rzeile @ 0\ mod _120 211 624 450 0_0. docx @ 11 4 @ @ 1

Pos: 51 / 200 Christ /37 0 RVC-BA (STANDARDM ODULE)/ 030 Sich erh eit/0 30- 006 0-0 050 Che misch e un d biol ogisch e Siche rheit @ 7\ mo d_13 093 310 408 18_ 68.d ocx @ 464 79 @ 3 @ 1

3.6.5 Chemical and biological safety

If pathogenic, toxic, or radioactive samples are intended to be evaporated,
it is in the responsibility of the user to ensure that all necessary safety
regulations, guidelines, precautions, and practices are adhered to
accordingly.

DANGER

• Infectious, toxic, pathogenic, and radioactive substances must be

evaporated in suitable accessories. Take suitable precautions for your
own safety!

• Do not evaporate corrosive products (especially substances containing

acids) without special protective measures and devices such as, for
example, a cooling trap for the protection of the vacuum pump. Contact
the manufacturer (see chapter 7.3 - "Service contact").

DANGER

• Special caution is necessary when handling azides, as a dangerous

explosive develops in combination with copper or nonferrous metals! It
is absolutely necessary to contact the manufacturer (see chapter 7.3 -
"Service contact").

• Keep informed about local measures to avoid harmful emissions

(depending on the substances to be evaporated).

• As personal protective equipment, safety gloves are required for the

use of the RVC.
The materials to be evaporated may, however, require additional
special safety measures (e.g. drying of infectious, toxic, radioactive, or
pathogenic substances)

Pos: 52 / 010 Unive rsalm odul e/ L ee rzeile @ 0\ mod _120 211 624 450 0_0. docx @ 11 4 @ @ 1

Pos: 53 / 200 Christ /37 0 RVC-BA (STANDARDM ODULE)/ 030 Sich erh eit/0 30- 006 0-0 060 Siche rheits hinweis e zu r Evap orati on @ 7\m od_ 130 933 104 119 3_68 .docx @ 4 648 7 @ 3 @ 1