7 Malfunctions and error correction

Version 04/2012, Rev. 1.4 of 25/11/2013 • sb

47

Translation of the original operating manual

Pos: 141 /20 0 Ch rist/3 70 RVC-BA (S TANDARDMODULE) /07 0 Stör ung en u nd Fehl ersuc he/ 070 Stör ung en u nd Fehl ersuc he === == == === == == === == == === == = @ 8\m od_ 1309 939 131 742 _68. docx @ 47 129 @ 1 @ 1

7 Malfunctions and error correction

Pos: 142 /20 0 Ch rist/3 70 RVC-BA (S TANDARDMODULE) /07 0 Stör ung en u nd Fehl ersuc he/ 070 -00 10 Allg emei ne Stö run gen @ 8\ mod _13 099 413 440 43_ 68.d ocx @ 472 24 @ 2 @ 1

7.1 General malfunctions

Malfunctions are displayed in the menu "Process & equipment info" (see
chapter 6.5.3.2 - "Process and equipment information"). A sound signal
sounds when an error message is generated.
• Eliminate the source of the problem (see the following chapters).
• Acknowledge the error messages by pressing the right-hand function

key.

Pos: 143 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1

Pos: 144 /20 0 Ch rist/3 71 RVC-BA (PROJEK TE)/RVC 2- 18 CD plus_ 2-1 8 CDplus _HCl/07 0 Stö run gen und Fe hlers uche /070 -00 10 -001 0 Fehl er meldu nge n RVC 2- 18_ 2-1 8 HCl @ 12\ mod _13 359 3986 448 7_6 8.do cx @ 6414 1 @ 3 @ 1

7.2 Error messages

Error

Possible reason

Correction

Lid cannot be opened

• The rotor chamber is not

aerated completely

• Aerate the rotor chamber

completely

Lid contacts error

• The system has detected an

open lid during the evaporation.
The rotor will be stopped
immediately. The lid must be

closed while the rotor is moving.

• Check the lid contacts. If

necessary, contact the after-
sales service (see chapter 7.3 -

"Service contact")

Rotor overtemperature

• The maximum permissible rotor

temperature has been

exceeded

• If the system does not cool

down, disconnect it immediately

from the power supply.

• Contact the after-sales service

(see chapter 7.3 - "Service

contact")

Chamber overtemperature

• The maximum permissible

chamber temperature has been

exceeded.

• If the system does not cool

down, disconnect it immediately

from the power supply.

• Contact the after-sales-service

(see chapter 7.3 - "Service

contact").

Permissible imbalance exceeded

• The rotor will be stopped

immediately, the process will

also be stopped.
– improper loading of the

rotor

– glass breakage during the

evaporation

– uneven evaporation

• Eliminate the imbalance and

restart the unit. If the error is still
present, contact the after sales-
service (see chapter 7.3 -

"Service contact")

VSK3000 calibration error

• Incorrect calibration of the

vacuum sensor VSK3000

• The sensor must be recalibrated

(see the separate operating

manual).