2 rotor chamber – Martin Christ RVC 2-18 CDplus User Manual

Page 54

Advertising
background image

RVC 2-18 CD& HCL

8 Maintenance and service

54

Version 04/2012, Rev. 1.4 of 25/11/2013 • sb

Translation of the original operating manual

Pos: 163 /20 0 Ch rist/3 71 RVC-BA (PROJEK TE)/RVC 2- 18 CD plus_ 2-1 8 CDplus _HCl/08 0 Wa rtun g un d In stan dhalt ung/ 080 -00 10- 002 1 Ro tork am mer HCl @ 1 2\m od_ 133 5341 579 279 _68. docx @ 64 031 @ 3 @ 1

8.1.2 Rotor chamber

If the rotor chamber is contaminated or if glass is broken, the rotor
chamber, the rotor shaft and the rotor must be cleaned immediately in order
to prevent corrosion and damage to the ball bearings.

Cleaning of the rotor chamber


1

Rotor chamber

2

Cap

3

Rotor shaft

Fig. 42: Rotor chamber with rotor shaft


• Open the lid.
• Switch off the mains power switch and disconnect the power cord.

DANGER

Hot surface!
Let the rotor chamber cool down to room temperature!


• Hold the rotor shaft and remove the cap by twisting it slightly and by

pulling it off upwards.

• Remove the hexagon socket screw with the supplied hexagon socket

key (size 5.5).

Advertising