RVC 2-18 CD& HCL

5 Set-up and connection

Version 04/2012, Rev. 1.4 of 25/11/2013 • sb

29

Translation of the original operating manual

Pos: 99 / 200 Christ /37 1 RVC-BA (PROJEKTE) /RVC 2-1 8 CDpl us_2 -18 CD plus_HCl /050 Aufstell ung und Anschluss /05 0-0 060 -00 30 PRO FESSIONAL: Ko nden satio n üb er eine Kü hlfalle mit Ve rteile rblock @ 1 2\mo d_1 335 260 055 746 _68. docx @ 64 007 @ 3 @ 1

5.5.3

PROFESSIONAL: Condensation of the vapours in a cooling trap with
distributor block

With a RVC 2-18 CD

plus

HCl and a cooling trap with distributor block

and glass cooling trap
The characteristic feature of a rotational vacuum concentrator (RVC) is the
combination of vacuum and rotation. During the process, a gas volume is
released. The released vapours that are carried off by the RVC condenser
in the glass cooling trap. This glass cooling trap is located inside a CT 02-
50 SR or CT 04-50 SR cooling trap.
The vacuum pump is connected to the distributor block. An electromagnetic
stop valve is connected to the RVC. The distributor block includes an
aeration valve. It ensures the aeration and facilitates the removal of the
glass cooling trap.
This method is suitable for aqueous and solvent-containing samples.


1 RVC
2 Vacuum hoses
3 Distributor block
4 Cooling trap
5 Vacuum pump
6 Electromagnetic stop

valve

Fig. 3: Combination of the RVC with a vacuum pump and cooling trap with a distributor block

front and rear view