8 Maintenance and service

Version 04/2012, Rev. 1.4 of 25/11/2013 • sb

53

Translation of the original operating manual

Pos: 157 /20 0 Ch rist/3 70 RVC-BA (S TANDARDMODULE) /08 0 Wart ung un d Inst and haltu ng/ 080 Wart ung und Inst and haltu ng= == == === == == === == == === == == = @ 8\mo d_1 309 944 760 465 _68. docx @ 47 238 @ 1 @ 1

8 Maintenance and service

The rotational vacuum concentrator and the accessories are subject to high
mechanical stress. Thorough maintenance performed by the user extends
the service life and prevents premature failure.

CAUTION

If corrosion or other damage occurs due to improper care, the
manufacturer cannot be held liable or subject to any warranty claims.


• Use soap water or other water-soluble, mild cleaning agents for

cleaning the RVC and the accessories.

• Do not use any corrosive or aggressive substances.
• Do not use solvents.
• Do not use agents with abrasive particles.
• Do not expose the RVC or its accessories to intensive UV radiation (e.g.

sunlight) or thermal stress (e.g. by heat generators).

Pos: 158 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1

Pos: 159 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1

Pos: 160 /20 0 Ch rist/3 70 RVC-BA (S TANDARDMODULE) /08 0 Wart ung un d Inst and haltu ng/ 080 -001 0 W artu ngsa rbei ten --- --- -- --- -- --- -- --- --- -- --- -- --- --- -- --- -- --- -- --- --- -- --- -- --- --- -- --- @ 8\ mo d_13 099 447 608 24_ 68.d ocx @ 472 46 @ 2 @ 1

8.1 Maintenance

Pos: 161 /20 0 Ch rist/3 70 RVC-BA (S TANDARDMODULE) /08 0 Wart ung un d Inst and haltu ng/ 080 -001 0- 0010 Rot ation s-Vaku um -Konze ntr ator @ 8\ mo d_1 3099 449 488 64_ 68.d ocx @ 472 54 @ 3 @ 1

8.1.1 Rotational vacuum concentrator

• Switch off the rotational vacuum concentrator via the mains power

switch and disconnect the power cord before cleaning.

• Carefully remove all product residues from the rotor chamber using a

cloth.

• If the rotational vacuum concentrator has been contaminated with toxic,

radioactive, or pathogenic substances, clean the rotor chamber
immediately with a suitable decontamination agent (depending on the
type of contamination).

WARNING

Take suitable precautions for your own safety if there is a risk of toxic,
radioactive, or pathogenic contamination.


• Check the safety guard of the fan (at the front of the bottom side) at

least once a year for dirt and clean it if necessary (e.g. with a vacuum
cleaner).

• Open the rotational vacuum concentrator when it is not in use so that

moisture can evaporate.