RVC 2-18 CD& HCL

5 Set-up and connection

Version 04/2012, Rev. 1.4 of 25/11/2013 • sb

25

Translation of the original operating manual

Pos: 81 / 200 Christ /37 0 RVC-BA (STANDARDM ODULE)/ 050 Au fstellu ng u nd Ans chluss/ 050 Aufst ellung un d Anschl uss= === == == === == == === == == == == @ 8\m od_ 130 933 583 2534 _68 .docx @ 4 658 1 @ 1 @ 1

5 Set-up and connection

Pos: 82 / 200 Christ /37 0 RVC-BA (STANDARDM ODULE)/ 050 Au fstellu ng u nd Ans chluss/ 050 -00 10 A ufstello rt, Eins atzo rt @ 8\m od_ 130 9335 833 534 _68 .docx @ 46 605 @ 2 @ 1

5.1 Installation site

Operate the RVC only in closed and dry rooms.
• The table must be stable and have a solid, even tabletop.
• Ensure sufficient ventilation. Do not place any paper, cloth, or similar

material behind or under the unit, since otherwise the air circulation will
be impaired.

• Keep a safety distance of at least 30 cm from the wall so that the vents

in the unit remain fully effective.

• The ambient temperature must be in the range of +10°C to +25°C.
• Do not subject the RVC to thermal stress, e.g. by positioning it near

heat generators.

• Avoid direct sunlight (UV radiation).

Pos: 83 / 010 Unive rsalm odul e/ L ee rzeile @ 0\ mod _120 211 624 450 0_0. docx @ 11 4 @ @ 1

Pos: 84 / 200 Christ /37 0 RVC-BA (STANDARDM ODULE)/ 050 Au fstellu ng u nd Ans chluss/ 050 -00 20 E ner gieve rsor gun g-- -- --- -- --- --- -- --- -- --- -- --- --- -- --- - @ 8\m od_ 1309 335 833 862 _68. docx @ 46 613 @ 2 @ 1

5.2 Power supply

Pos: 85 / 200 Christ /37 1 RVC-BA (PROJEKTE) /RVC 2-1 8 CDpl us_2 -18 CD plus_HCl /050 Aufstell ung und Anschluss /05 0-0 020 -00 11 An schluss art RVC 2-1 8 HCl @ 12\ mod _133 697 526 006 4_6 8.doc x @ 6 446 3 @ 3 @ 1

5.2.1 Connection

DANGER

The operating voltage on the name plate must correspond to the local
supply voltage


CHRIST rotational vacuum concentrators are units of safety class I.Units of
the type RVC 2-18 CD

plus

HCl have a three-wire power cord with an IEC

C13 connector (see chapter 10 - "Technical data").
An equipotential bonding screw is located on the back (see chapter 2.1.1 -
"Functional and operating elements"). This equipotential bonding screw
can be used to perform an earth conductor check.

Pos: 86 / 010 Unive rsalm odul e/ L ee rzeile @ 0\ mod _120 211 624 450 0_0. docx @ 11 4 @ @ 1

Pos: 87 / 200 Christ /37 0 RVC-BA (STANDARDM ODULE)/ 050 Au fstellu ng u nd Ans chluss/ 050 -00 20- 002 0 Sich eru nge n ba useits @ 8\ mod _13 093 3583 459 6_6 8.do cx @ 4662 9 @ 3 @ 1

5.2.2 Customer-provided fuses

Typically, the RVC must be protected with 16 Amp G fuses that are to be
provided by the customer.

Pos: 88 / 010 Unive rsalm odul e/ Seit enwec hsel @ 0\m od_ 120 211 6244 312 _0. docx @ 10 5 @ @ 1