Martin Christ RVC 2-18 CDplus User Manual

Page 30

Advertising
background image

RVC 2-18 CD& HCL

5 Set-up and connection

30

Version 04/2012, Rev. 1.4 of 25/11/2013 • sb

Translation of the original operating manual

Pos: 101 /20 0 Ch rist/3 71 RVC-BA (PROJEK TE)/RVC 2- 18 CD plus_ 2-1 8 CDplus _HCl/05 0 Aufst ellun g un d Anschl uss/0 50- 006 0-0 040 PROFESSIONAL TWIN: Ko nde nsatio n üb er eine Kü hlfalle mit zw ei RVC HCl @ 1 2\m od_ 133 533 260 035 7_68 .docx @ 6 402 1 @ 3 @ 1

5.5.4

PROFESSIONAL TWIN: Condensation of the vapours in a cooling trap with
distributor block and two RVCs

With two RVC 2-18 CD

plus

HCl and a cooling trap with distributor block

and glass cooling trap
The characteristic feature of a rotational vacuum concentrator (RVC) is the
combination of vacuum and rotation. During the process, a gas volume is
released. The released vapours that are carried off by the RVC condenser
in the glass cooling trap. This glass cooling trap is located inside a CT 02-
50 SR or CT 04-50 SR cooling trap.
The vacuum pump is connected to the distributor block. An electromagnetic
stop valve is connected to each of the RVCs. The distributor block includes
an aeration valve. It ensures the aeration and facilitates the removal of the
glass cooling trap.
The method is suitable for aqueous and solvent-containing samples.


1 RVC
2 Vacuum pump
3 Vacuum hoses
4 Distributor block
5 Cooling trap
6 Electromagnetic stop

valve

Fig. 4: Combination of two RVCs with a vacuum pump and cooling trap

with a distributor block, front and rear view

Pos: 102 /01 0 Univ ersal mod ule/ Sei tenwe chsel @ 0\ mod _12 021 1624 431 2_0 .docx @ 1 05 @ @ 1