7 Malfunctions and error correction

48

Version 04/2012, Rev. 1.4 of 25/11/2013 • sb

Translation of the original operating manual

Pos: 146 /20 0 Ch rist/3 71 RVC-BA (PROJEK TE)/RVC 2- 18 CD plus_ 2-1 8 CDplus _HCl/07 0 Stö run gen und Fe hlers uche /070 -00 10 -002 0 Pr ozess meld ung en RVC 2 -18 _2- 18 HCl @ 12\ mo d_1 3359 398 649 60_ 68.d ocx @ 641 49 @ 3 @ 1

7.2.1 Process messages

Error

Possible reason

Correction

Mains power failure

• A mains power failure will be

detected if the power supply is
interrupted under normal

conditions (run mode).

• The process will continue once

the mains power is available
again. The process times will be
reset to zero. The set control

values will be preserved.

Factory settings loaded

• All parameters are reset to the

delivery status of the system.

• The factory settings will be

loaded if the corresponding
menu command is executed in
the service menu or if an error
occurred in the parameter

memory.

Nominal rotor speed not reached

• After two minutes, the rotor has

still not reached 90% of its

nominal speed.
– set speed is too high
– drive moves sluggishly

• Check the set speed.
• When the rotor is at a standstill

and the system is switched off,
check whether the drive moves
sluggishly, by hand. Contact the
after-sales service (see chapter

7.3 - "Service contact").

Safety pressure not reached

• The safety pressure inside the

rotor chamber must be reached
within 5 minutes after the start

of an evaporation process.

• Check whether the value that

was entered for the safety
pressure is not too low.

• Check the components for any

possible leaks.

Pos: 147 /01 0 Univ ersal mod ule/ Lee rzeile @ 0\ mo d_12 021 162 445 00_ 0.doc x @ 1 14 @ @ 1

Pos: 148 /20 0 Ch rist/3 70 RVC-BA (S TANDARDMODULE) /07 0 Stör ung en u nd Fehl ersuc he/ 070 -00 10- 003 0 Syste mm eldu ngen @ 8 \mo d_1 309 9391 353 52_ 68.d ocx @ 472 01 @ 3 @ 1

7.2.2 System messages

Error

Possible reason

Correction

Invalid system type

• Contact the after-sales service

(see chapter 7.3 - "Service

contact").

IO communication error 0x21

• Failure of the CDplus IO module • Switch the system off and then

on again. If the error occurs
again, contact the after-sales
service (see chapter 7.3 -

"Service contact")

IO communication error 0x30

• Failure of the reading unit of the

electronic rotor system

• Switch the system off and then

on again. If the error occurs
again, contact the after-sales
service (see chapter 7.3 -

"Service contact")

IO communication error 0x40

• Failure of the motor control

system

• Switch the system off and then

on again. If the error occurs
again, contact the after-sales
service (see chapter 7.3 -

"Service contact")

IO communication error 0x41

Failure of the imbalance
detection module

• Switch the system off and then

on again. If the error occurs
again, contact the after-sales
service (see chapter 7.3 -

"Service contact")