Installation and connections, 1 electrical connection, Caution – BINDER MK 53 User Manual

Page 21

Advertising
background image

MK (E2.1) 10/2014

page 21/73

4.

Installation and connections

4.1 Electrical connection

The alternating climate chamber MK is supplied ready for connection.

Shockproof plug, power supply voltage 230 V (1N~) +/- 10 %, 50 Hz

Fixed power connection cable 1800 mm / 5.9 ft in length

The socket must also provide a protective conductor.

Prior to connection and start-up, check the power supply voltage. Compare the values to the specified

data located on the unit’s type plate (unit front behind the door, bottom left-hand, chap. 1.4)

When connecting, please observe the regulations specified by the local electricity supply company and

as well as the VDE directives (for Germany). We recommend the use of a residual current circuit

breaker.

Pollution degree (acc. to IEC 61010-1): 2

Over-voltage category (acc. to IEC 61010-1): II

CAUTION

Danger of incorrect power supply voltage.
Damage to the equipment.

Check the power supply voltage before connection and start-up.

Compare the power supply voltage with the data indicated on the type plate.

See also electrical data (chap. 18.4).

To completely separate the unit from the power supply, you must disconnect the power plug.

Install the unit in a way that the power plug is easily accessible and can be easily pulled in case

of danger.

Advertising