page 28/73

6.4 Menu settings in the “User Level” menu

User Level

Date and time

Summer time

Language

Temperature unit

Buzzer
Safety controller

User-code No.

English

°C

1

Active


Date and time
Enter the actual date and time to provide the proper measurement records. Data is displayed in the chart

recorder function (chap. 7) of the controller and will remain stored in case of a power failure.

Summer time
Time is set one hour in advance during the summer time period.
Setting the summer time switch:

Off: No change to summer time occurs

User timed: Beginning and end of summer time can be set individually

Automatic: The summer time arrangement for central Europe is enabled

(summer time from last Sunday of March until last Sunday of October)

Language
Select the menu language as German, English, or French (chap. 6.1).

Temperature unit

Do NOT change the temperature unit from °C to °F.

Buzzer
Audible alarm buzzer

Inactive: No audible alarm will sound if an alarm event happens (chap. 12).

Active: An audible alarm will sound in case of an alarm event (chap. 12).

Safety controller
Enter a safety controller tolerance limit to prevent temperature from exceeding this setting. For setting,

see chap. 11.3.

User-Code No.
Change the password (“user code”) needed to access the “User settings” menu. Factory default setting

+00001.

Keep in mind any change of the user password. There is no entrance to this menu without the

correct password.