Performance after completing the program:
The controller changes to Idle Mode. The heating and the cooling are inactive; the chamber approximates

ambient temperature. The fan is off.

9.2 Selecting between set-point ramp and set-point step

Temperature set-points always refer to the start of a program section, i.e., at the beginning of each pro-

gram section the entered temperature set-point is targeted. During program section operation, the tem-

perature gradually passes to the set-point entered for the next section.
By appropriate planning of the program section timing, you can enter all kinds of temperature transitions.

Gradual temperature changes “set-point ramp”
The set-point changes its value gradually while proceeding from one program section to the next one

during the programmed section length. The actual temperature value (X) follows the continually moving

set-point (W) at any time.

Program sections with constant temperature
The initial values of two subsequent program sections are identical; so the temperature remains con-

stant during the whole time of the first program section.

Sudden temperature changes “set-point step”
Steps are temperature changes (ramps) that occur during a very short interval. A section with a differ-

ent set-point follows two program sections with an identical set-point. If the duration of this transitional

program section is very short (minimum entry 1 sec), the temperature change will proceed rapidly with-

in the minimum amount of time.

01

02

03

04

05

06

07

08

09

W

t

Figure 8: Possible temperature transitions

The following chapter offers examples of programming a set-point ramp and a set-point step.

9.3 Program entry as set-point ramp or as set-point step

In order to avoid incorrect programming, we recommend plotting the temperature profile (chart template in

chap. 0) and entering the values into a table (template in chap. 9.10).
The controller provides 8 operation lines that can be activated or de-activated for each program section.

Operating contact 1 serves to control the bedew protection (chap. 10). The other operation lines are non-

functional.