12. Notification and alarm functions

12.1 Notification and alarm system overview (auto diagnosis system)

The alternating climate chamber MK provides notification and alarm functions.

Visual indications of notifications or error message are blue notes on the display of the MB1 controller.

Visual indications of an alarm message are red notes with an alarm bell symbol.

In addition, there is an audible alert, if you did not deactivate the buzzer in the “User level” menu (chap.

6.4).

Event

Note

(blue field)

Alarm

(red field)

Door open

DOOR OPEN

immediately

Operation line 1 (bedew protection) activated

DRY

immediately

Limit value of safety controller exceeded

TEMPERATURE LIMIT

immediately

Power failure

With option over/under temperature safety device class 2 (chap. 11.3):

Exceeding the maximum / minimum tempera-

ture

TEMP CLASS

immediately

With option keyboard locking (chap. 14.3):

Keyboard locked

KEY LOCK

immediately

The indicated intervals refer to the time after occurrence of the error or notified condition.

12.2 Resetting the notifications or alarm messages

The “RESET” button, which serves to acknowledge and reset the indication, will become visible automati-

cally whenever a notification or an alarm message appears.
1. Depending on the type of error, remove the cause of the disturbance or wait until the unit compensates

for the reason of the error.

2. Press the “RESET” button to reset the notification or alarm message.

CAUTION

In case the “RESET” button does not cancel the notification or alarm indication, the reason

for the disturbance was not removed correctly

Contact BINDER service.