MK (E2.1) 10/2014

page 37/73

Program entry as set-point ramp (example)

W/°C

t/min.

01

02

03

04

-40

-20

0

20

40

60

30

120

180

380

05

Minimum

tolerance

Maximum

tolerance

Operation line 1 = Bedew protection

ON

OFF

t/min.

Program table corresponding to the diagram above:

Program

section

Set-point

temp.

Fan

Section

time

Operation

line1

Target

section

No. of

cycles

Min.

tolerance

Max.

tolerance

01

-20

100 %

00:30:00

Off

1

0

-1999

+9999

02

0

100 %

01:30:00

Off

1

0

-5

+5

03

60

100 %

01:00:00

Off

1

0

-2

+2

04

60

100 %

03:20:00

Off

1

0

-1999

+9999

05

-40

100 %

00:00:01

Off

1

0

-1999

+9999

Now enter the values of the above program table into one of the 25 program places of the controller MB1: