14. Options

14.1 Communication software APT-COM™ 3 DataControlSystem (option)

The alternating climate chamber MK is regularly equipped with a serial interface RS 422 that can connect

the BINDER communication software APT-COM™ 3 DataControlSystem. The actual temperature values

are given at adjustable intervals. Programming can be performed graphically via PC. Up to 30 chambers

with RS 422 interface can be cross-linked. For further information, please refer to the operating manual of

the BINDER communication software APT-COM™.
Pin allocation of the RS 422 interface:

pin 2: RxD (+)
pin 3: TxD (+)
pin 4: RxD (-)
pin 5: TxD (-)

pin 7: Ground

14.2 Analog output for temperature (option)

With this option the chamber is equipped with an analog output 4-20 mA for temperature. This output

permits transmitting data to external data registration systems or devices.
The connection is realized as a DIN socket (8) in the lateral control panel as follows:

ANALOG OUTPUT 4-20 mA DC
PIN 1: Temperature –

PIN 2: Temperature +
Temperature range: -40 °C / -40 °F up to +180 °C / -356 °F
A suitable DIN plug is enclosed.

Figure 9: Pin allocation of the DIN socket (8) for option analog output

14.3 Keyboard locking (option)

The keyboard of the MB1 controller can be locked and unlocked via the key switch (option). In locked

state, no entries to the controller are possible.

Locked keyboard:

Switch position vertical

Unlocked keyboard: Switch position to the right

You can remove the key only when the keyboard is locked.





Key switch

Figure 10: Keyboard locking (option)

If the keyboard is locked, the notification “KEY LOCK” is dis-

played on the controller MB1 display next to a flashing blue in-

formation symbol.

HAND

W

VIEW->

KEY LOCK